Poseidon

Pressure and level transmitters

A submersible sensor on a cable. It reads the pressure of the liquid standing above it and reports the level as a 4–20 mA signal. Series 3000 pairs the sensor with a separate amplifier unit and is approved for hazardous areas; series 4000 is one piece.

Ask about Poseidon

Schematic drawing, not to scale.

Specifications

Measuring range, gauge
3000, 3300: up to 35 bar (smallest span 0.15 bar). 4000, 4300: up to 50 bar (smallest span 0.25 bar)
Measuring range, absolute
3000, 3300: up to 35 bar (smallest span 0.5 bar). 4000, 4300: up to 50 bar (smallest span 1 bar)
Negative pressure
down to −1 bar
Accuracy
3000, 3300: ±0.25 % FS at 20 °C, total error band ±0.1 % FS per °C. 4000, 4300: ±0.05 % FS
Output
two-wire 4–20 mA current loop
Supply
3000, 3300: 10 to 28 V DC. 4000, 4300: 11 to 30 V DC. Protected against reverse voltage
Media temperature
3000, 3300: −20 to +85 °C. 4000, 4300: −40 to +135 °C
Sensor
3000, 3300: AISI 316L housing and diaphragm, silicone oil filling. 4000, 4300: AISI 316L housing, ceramic diaphragm (Al₂O₃ 96 %)
Cable
PVC, to the length you ask for; up to 100 m with ventilation tube on the 3000 and 3300
Process connection
G ¼" to 1", filter, flange (DIN, JIS) or M20×1.5 EN 837 depending on the model; M21×1.5 adaptor with fixing gland on the 3300 and 4300
Protection class
IP68; amplifier unit of the 3000 and 3300: IP66
Approvals, 3000 and 3300
CE marked. IECEx and ATEX, intrinsic safety: Ex ia IIC T4 Ga, II 1 G (IECEx EMT 17.0008X)
Documents
calibration report with every instrument

The models

Where it works

  • Ballast tanks
  • Fuel and oil tanks
  • Fresh water tanks
  • Marine and offshore
  • Process industry
  • Environmental engineering

Options and versions

  • Gauge or absolute reference
  • Cable to the depth of your tank
  • Thread, filter, flange or adaptor with fixing gland
  • Intrinsically safe system (3000, 3300)

The rest of the range